Introduction
Capable of clean room wall mount, the BFD-10S pilot freeze dryer with clean room configuration perfectly fit laboratories and pilot plants that require a higher hygiene level.
The machine is designed for freeze drying products that are loaded in vials or bulk.
Technical Parameters
Model | BFD-10S Clean Room Config |
---|---|
Shelves Area | 0.47~1.1m2 |
Shelves No. | (3~7)+1 |
Shelves Size | 350*450mm |
Shelves Temp. | -55C~+70C |
Condenser | 15Kgs/24Hrs, 30kgs total. |
Condenser Temp. | -85C |
Final Vacuum | 0.01mBar |
Power Install | 8Kw |
Size | W/1000*D/1800*H/1950mm |
Weight | 800Kgs |
Vials 2R/4R | 4452 |
Vials 6R/8R | 2233 |
Vials 10R/15R | 1890 |
Technical Description
- All-include body, with casters and height adjustable feet.
- Inbuilt vacuum pump with oil mist filter.
- Advanced PLC and user-friendly HMI based control system, LYO-PILOT software.
- Pressure vessel standard S/S drying chamber.
- Transparent acrylic chamber door.
- Cooling and heating shelves from -55C to +70C.
- Hydraulic stoppering for vials closing, top to bottom.
- Horizontal cylindrical ice condenser, lowest -85C.
- Transparent acrylic condenser door.
- Condenser auto defrost.
- 2 stage oil sealed rotary vane vacuum pump.
- Pirani type vacuum sensor.
- Vacuum final 0.01mbar and is adjustable.
- Air cooled, cascade compressor system.
- Documents and validation package include in standard offer.
System Performance
- Chamber inner wall roughness Ra<0.6um
- Shelves flatness <0.5mm/m
- Shelves roughness Ra<0.4um
- Shelf cooling: +20C ~ -40C <60min
- Cooling final: -55C (+/-2C)
- Shelf heating: -40C ~ +20C <60min
- Heating final: +70C (+/-1.5C)
- Condenser inner wall roughness Ra<0.6um
- Condenser coils roughness Ra<0.4um
- Condenser cooling: RT ~ -40C <30min
- Condenser final: -85C
- Defrosting: <60Min
- Vacuum final: <1Pa