General Purpose Freeze Dryer

Commercial Freeze Dryer for Various Industries

Vekuma’s FD-20L freeze dryer is a high performance commercial freeze dryer that designed for a variety of industrial applications. Typical applications include bacterial, yogurt, nutrient and probiotics lyophilization. The machine is designed to process bulk materials, products are loaded and freeze dried in material trays.

The FDL general purpose freeze dryers feature rectangular drying chamber which means highest space using rate and exceptional temperature uniformity. The proven advantages result they are particularly fit freeze dry heat sensitive items and guarantee the consistently freeze dried results.

Technical Specifications

Model FD-20L
Shelf Area 19.8m2
Useable Shelves 11
Shelves Size 1200*1500mm
Shelves Temp. -50C~+70C
Condenser 200Kgs/24Hrs, 400Kgs Max
Condenser Temp. -75C
Final Vacuum <5Pa
Power Install 68Kw
Size W/2000*L/7500*H/2800mm

Technical Description

  • All-in-one design, easy transport and install.
  • Plug and play deliver.
  • Advanced PLC and user-friendly HMI based control system, FD-MEGA V2.0 interface.
  • Entirely S/S304 fabricated include chamber, shelves, trays and condenser.
  • Rectangular drying chamber, high space use rate and high temperature uniformity.
  • Heavy duty hinged door with large sight glass.
  • Cooling and heating shelves temperature from -50C to +70C.
  • External type ice condenser, lowest -75C.
  • Condenser hot water defrost.
  • Vacuum system adopt 2 stage oil sealed pumps + roots pump.
  • System final vacuum <0.05mbar, vacuum adjustable.
  • Pirani type vacuum sensor.
  • Water cooled, semi-hermetic compressor system.
  • Documents and validation package include in standard offer.

System Performance

  • Shelf cooling: +20C ~ -40C <60min
  • Shelves lowest: -50C
  • Shelf heating: -40C ~ +20C <60min
  • Shelves highest: +70C
  • Shelf temperature uniformity: <+/-1C
  • Condenser cooling: RT ~ -40C <30min
  • Condenser final: -75C
  • Defrosting: <60Min (+65C Water)
  • Vacuumize to 100Pa: <30Min
  • Vacuum final: <1Pa
  • Chamber leak rate <5Pa L/Sec
  • Chamber inner wall roughness Ra<0.6um.
  • Shelves flatness <0.5mm/m.
  • Shelves roughness Ra<0.4um.
  • Condenser inner wall roughness Ra<0.6um.
  • Condenser coils roughness Ra<0.4um.
Share This Post